Validation of bohr dead space measured by volumetric capnography

Nenhuma Miniatura disponível
Citações na Scopus
1
Tipo de produção
bookPart
Data de publicação
2012
Título da Revista
ISSN da Revista
Título do Volume
Editora
SPRINGER BERLIN HEIDELBERG
Autores
TUSMAN, G.
SUAREZ-SIPMANN, F.
HEDENSTIERNA, G.
BOHM, S. H.
Citação
Tusman, G.; Suarez-Sipmann, F.; Borges, J. B.; Hedenstierna, G.; Bohm, S. H.. Validation of bohr dead space measured by volumetric capnography. In: . APPLIED PHYSIOLOGY IN INTENSIVE CARE MEDICINE 1: PHYSIOLOGICAL NOTES - TECHNICAL NOTES - SEMINAL STUDIES IN INTENSIVE CARE, THIRD EDITION: SPRINGER BERLIN HEIDELBERG, 2012. p.195-199.
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
Purpose: Bohr’s dead space (VDBohr) is commonly calculated using end-tidal CO2 instead of the true alveolar partial pressure of CO2 (PACO2). The aim of this work was to validate VDBohr using PACO2 derived from volumetric capnography (VC) against VDBohr with PACO2 values obtained from the standard alveolar air formula. Methods: Expired gases of seven lung-lavaged pigs were analyzed at different lung conditions using mainstream VC and multiple inert gas elimination technique (MIGET). Paco2 was determined by VC as the midpoint of the slope of phase III of the capnogram, while mean expired partial pressure of CO2 (PeCO2) was calculated as the mean expired fraction of CO2 times the barometric minus the water vapor pressure. Miget estimated expired CO2 output (VCO2) and PeCO2 by its V/Q algorithms. Then, PACO2 was obtained applying the alveolar air formula (PACO2 = VCO2/alveolar ventilation). Results: We found close linear correlations between the two methods for calculating both PACO2 (r = 0.99) and VDBohr (r = 0.96), respectively (both p\0.0001). Mean PACO2 from VC was very similar to the one obtained by MIGET with a mean bias of -0.10 mmHg and limits of agreement between -2.18 and 1.98 mmHg. Mean VDBohr from VC was close to the value obtained by MIGET with a mean bias of 0.010 ml and limits of agreement between -0.044 and 0.064 ml. Conclusions: VDBohr can be calculated with accuracy using volumetric capnography. © Springer-Verlag Berlin Heidelberg 2006, 2009, 2012.
Palavras-chave
Bohr’s dead space, Carbon dioxide, Gas exchange, PACO 2, Tidal volume, Volumetric capnography
Referências
  1. Bohr, C., Über die lungeatmung (1891) Skand Arch Physiol, 2, pp. 236-238
  2. Haldane, J.S., Priestley, J.G., The regulation of the lung-ventilation (1905) J Physiol, 32, pp. 225-266
  3. Krogh, A., Lindhard, J., On the average composition of the alveolar air and its variations during the respiratory cycle (1913) J Physiol, 47, pp. 431-445
  4. Fowler, W.S., Lung function studies ii. The respiratory dead space (1948) Am J Physiol, 154, pp. 405-416
  5. Aitken, R.S., Clark-Kennedy, A.E., On the fluctuation in the composition of the alveolar air during the respiratory cycle in muscular excercise (1928) J Physiol, 65, pp. 389-411
  6. Fletcher, R., Jonson, B., The concept of deadspace with special reference to the single breath test for carbon dioxide (1981) Br J Anaesth, 53, pp. 77-88
  7. Breen, P.H., Mazumdar, B., Skinner, S.C., Comparison of end-tidal pco2 and average alveolar expired pco2 during positive end-expiratory pressure (1996) Anesth Analg, 82, pp. 368-373
  8. Tusman, G., Scandurra, A., Bohm, S.H., Suarez Sipmann, F., Clara, F., Model fitting of volumetric capnograms improves calculations of airway dead space and slope of phase iii (2009) J Clin Monit Comput, 23, pp. 197-206
  9. Kallet, R.H., Daniel, B.M., Garcia, O., Matthay, M.A., Accuracy of physiologic dead space measurements in patients with ards using volumetric capnography: Comparison with the metabolic monitor method (2005) Respir Care, 50, pp. 462-467
  10. West, J.B., Ventilation-perfusion inequality and overall gas exchange in computer models of the lung (1969) Respiration Physiol, 7, pp. 88-110
  11. Wagner, P.D., Saltzman, H.A., West, J.B., Measurement of continuous distributions of ventilation-perfusion ratios: Theory (1974) J Appl Physiol, 36, pp. 588-599
  12. Dubois, A.B., Britt, A.G., Fenn, W.O., Alveolar co2 during the respiratory cycle (1951) J Appl Physiol, 4, pp. 535-548
  13. Jordanoglou, J., Koulouris, N., Kyroussis, D., Rapakoulias, P., Vassalos, P., Madianos, J., Measurement of effective alveolar carbon dioxide tension during spontaneous breathing in normal subjects and patients with chronic airways obstruction (1995) Thorax, 50, pp. 240-244
  14. Enghoff, H., Volumen inefficax. Bemerkungen zur frage des schadlichen raumes (1938) Uppsala Lakareforen Forhandl, 44, pp. 191-218
  15. Riley, R.L., Cournand, A., Ideal alveolar air and the analysis of ventilation-perfusion relationships in the lungs (1949) J Applied Physiol, 1, pp. 825-847
  16. Suter, P.M., Fairley, H.B., Isenberg, M.D., Optimum end-expiratory airway pressure in patients with acute pulmonary failure (1975) N Engl J Med, 292, pp. 284-289
  17. Hedenstierna, G., Sandhagen, B., Assessing dead space. A meaningful variable? (2006) Minerva Anestesiol, 72, pp. 521-528