Please use this identifier to cite or link to this item: https://observatorio.fm.usp.br/handle/OPI/29475
Title: Evaluation of adenosine triphosphate test for cleaning assessment of gastroscopes and the effect on workload in a busy endoscopy center
Authors: SCHMITT, CristianeMACIEL, Amanda Luiz PiresBOSZCZOWSKI, IcaroSILVA, Thais Pereira daNEVES, Eliane Aparecida JobROSSINI, Giulio FabioRIZEK, CamilaCOSTA, Silvia FigueiredoLOURENCO, Rogerio FerreiraALFA, Michelle J.
Citation: AMERICAN JOURNAL OF INFECTION CONTROL, v.46, n.10, p.1110-1114, 2018
Abstract: Objective: Using adenosine triphosphate (ATP) tests to assess manual cleaning of gastroscopes and to determine the associated workload in a busy endoscopy unit. Methods: Patient-used gastroscopeswere sampled before and after cleaning to assess ATP levels, bioburden, and protein. Samples were collected by flushing 20 mL of sterile water through the biopsy port to the distal end. Time spent for reprocessing and performing the ATP test was recorded. Results: Twenty-four samples were collected from 10 gastroscopes. After manual cleaning, 14/24 (58.3%) samples had no microbial growth (mean, 21 colony-forming units/cm(2)), and in 22/24 (91.7%) samples the protein was undetectable (mean, 0.04 mu g/cm(2)). ATP test was above the cutoff (200 relative light units [RLU]) in 17/24 (70.8%) samples (mean, 498 RLU). After the second cleaning, 11/17 (64.7%) gastroscopes still failed the ATP test (mean, 321.2 RLU). The mean time spent to perform manual cleaning and ATP tests was 16 and 8 minutes, respectively. Hence, each test increased the length of time for cleaning plus testing cleanliness by 50%. Conclusion: Further studies regarding the optimal cutoff for ATP tests are needed. ATP tests for cleaning monitoring are easy to perform and provide immediate feedback to the team. However, the increased workload needs to be considered.
Appears in Collections:

Artigos e Materiais de Revistas Científicas - FM/MIP
Departamento de Moléstias Infecciosas e Parasitárias - FM/MIP

Artigos e Materiais de Revistas Científicas - HC/ICESP
Instituto do Câncer do Estado de São Paulo - HC/ICESP

Artigos e Materiais de Revistas Científicas - HC/ICHC
Instituto Central - HC/ICHC

Artigos e Materiais de Revistas Científicas - LIM/54
LIM/54 - Laboratório de Bacteriologia


Files in This Item:
File Description SizeFormat 
art_SCHMITT_Evaluation_of_adenosine_triphosphate_test_for_cleaning_assessment_2018.PDF
  Restricted Access
publishedVersion (English)375.3 kBAdobe PDFView/Open Request a copy

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.