Please use this identifier to cite or link to this item: https://observatorio.fm.usp.br/handle/OPI/30678
Title: Modeling, characterization and test of a pediatric ventricular assist device
Authors: SOUSA, D. L.CORDEIRO, T. D.MELO, T. R.ROCHA NETO, J. S. daCESTARI, I. A.LIMA, A. M. N.
Citation: 2017 JOINT IMEKO TC1-TC7-TC13 SYMPOSIUM: MEASUREMENT SCIENCE CHALLENGES IN NATURAL AND SOCIAL SCIENCES, v.1044, article ID UNSP 012047, 6p, 2018
Abstract: This paper presents a OD model for the pulsatile pediatric Ventricular Assist Device (pVAD), as well as its characterization and test results obtained by using an experimental hydraulic system that provides operational conditions close to the physiological conditions observed in pediatric patients. The measurements of the physical quantities used in characterization of the pVAD have been obtained by using the LabVIEW platform. The air chamber pressure signal is used to model the pneumatic actuation system behaviour, whereas the pressure-volume relationship inside the blood chamber is applied to model the pVAD compliance. The set of test and validation results demonstrated the correctness of the proposed model.
Appears in Collections:

Comunicações em Eventos - HC/InCor
Instituto do Coração - HC/InCor


Files in This Item:
File Description SizeFormat 
art_SOUSA_Modeling_characterization_and_test_of_a_pediatric_ventricular_2018.PDFpublishedVersion (English)1.43 MBAdobe PDFThumbnail