Please use this identifier to cite or link to this item: https://observatorio.fm.usp.br/handle/OPI/18663
Title: Controlling the cardiovascular response under synchronized ventricular assist device support
Authors: CORDEIRO, Thiago D.SOUSA, Daniel L.CESTARI, Idagene A.LIMA, Antonio M. N.
Citation: 2016 IEEE INTERNATIONAL CONFERENCE ON AUTOMATICA (ICA-ACCA), 2016
Abstract: Ventricular assist devices are mechanical blood pumps used to provide circulatory support in patients waiting for a heart transplantation. In many cases, when the support lasts for months it may be advantageous to use a pump with a physiological control system to automatically adjust pump output in response to changes in the cardiovascular system (CVS). In this paper, we present a physiological controller for a pulsatile ventricular assist device (PVAD) that controls the device ejection volume considering the heart beat frequency. A mathematical model of the CVS based on analog electric circuits is used to analyze the cardiovascular response under ventricular assistance. The physiological controller is shown to be able to improve simulated hemodynamic conditions.
Appears in Collections:

Comunicações em Eventos - HC/InCor
Instituto do Coração - HC/InCor


Files in This Item:
File Description SizeFormat 
art_CORDEIRO_Controlling_the_cardiovascular_response_under_synchronized_ventricular_assist_2016.PDF
  Restricted Access
publishedVersion (English)1.68 MBAdobe PDFView/Open Request a copy

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.