Please use this identifier to cite or link to this item: https://observatorio.fm.usp.br/handle/OPI/31675
Title: Efficacy, Safety, and Performance of Isolated Left vs. Right Ventricular Pacing in Patients with Bradyarrhythmias: A Randomized Controlled Trial
Authors: CREVELARI, Elizabeth SartoriSILVA, Katia Regina daALBERTINI, Caio Marcos de MoraesVIEIRA, Marcelo Luiz CamposMARTINELLI FILHO, MartinoCOSTA, Roberto
Citation: ARQUIVOS BRASILEIROS DE CARDIOLOGIA, v.112, n.4, p.410-420, 2019
Abstract: Background: Considering the potential deleterious effects of right ventricular (RV) pacing, the hypothesis of this study is that isolated left ventricular (LV) pacing through the coronary sinus is safe and may provide better clinical and echocardiographic benefits to patients with bradyarrhythmias and normal ventricular function requiring heart rate correction alone. Objective: To assess the safety, efficacy, and effects of LV pacing using an active-fixation coronary sinus lead in comparison with RV pacing, in patients eligible for conventional pacemaker (PM) implantation. Methods: Randomized, controlled, and single-blinded clinical trial in adult patients submitted to PM implantation due to bradyarrhythmias and systolic ventricular function >= 0.40. Randomization (RV vs. LV) occurred before PM implantation. The main results of the study were procedural success, safety, and efficacy. Secondary results were clinical and echocardiographic changes. Chi-squared test, Fisher's exact test and Student's t-test were used, considering a significance level of 5%. Results: From June 2012 to January 2014, 91 patients were included, 36 in the RV Group and 55 in the LV Group. Baseline characteristics of patients in both groups were similar. PM implantation was performed successfully and without any complications in all patients in the RV group. Of the 55 patients initially allocated into the LV group, active-fixation coronary sinus lead implantation was not possible in 20 (36.4%) patients. The most frequent complication was phrenic nerve stimulation, detected in 9 (25.7%) patients in the LV group. During the follow-up period, there were no hospitalizations due to heart failure. Reductions of more than 10% in left ventricular ejection fraction were observed in 23.5% of patients in the RV group and 20.6% of those in the LV group (p = 0.767). Tissue Doppler analysis showed that 91.2% of subjects in the RV group and 68.8% of those in the LV group had interventricular dyssynchrony (p = 0.022). Conclusion: The procedural success rate of LV implant was low, and the safety of the procedure was influenced mainly by the high rate of phrenic nerve stimulation in the postoperative period.
Appears in Collections:

Artigos e Materiais de Revistas Científicas - FM/MCP
Departamento de Cardio-Pneumologia - FM/MCP

Artigos e Materiais de Revistas Científicas - HC/ICESP
Instituto do Câncer do Estado de São Paulo - HC/ICESP

Artigos e Materiais de Revistas Científicas - HC/InCor
Instituto do Coração - HC/InCor

Artigos e Materiais de Revistas Científicas - LIM/11
LIM/11 - Laboratório de Cirurgia Cardiovascular e Fisiopatologia da Circulação

Artigos e Materiais de Revistas Científicas - ODS/03
ODS/03 - Saúde e bem-estar


Files in This Item:
File Description SizeFormat 
art_CREVELARI_Efficacy_Safety_and_Performance_of_Isolated_Left_vs_2019_eng.PDFpublishedVersion (English)360.81 kBAdobe PDFThumbnail
View/Open
art_CREVELARI_Efficacy_Safety_and_Performance_of_Isolated_Left_vs_2019_por.PDFpublishedVersion (Portuguese)428.51 kBAdobe PDFThumbnail
View/Open

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.