dc.description.abstractCochlear implantation is a safe and reliable method for auditory restoration in patients with severe to profound hearing loss. Objective: To describe the surgical complications of cochlear implantation. Materials and Methods: Information from 591 consecutive multichannel cochlear implant surgeries were retrospectively analyzed. All patients were followed-up for at least one year. Forty-one patients were excluded because of missing data, follow-up loss or middle fossa approach. Results: Of 550 cochlear implantation analyzed, 341 were performed in children or adolescents, and 209 in adults. The mean hearing loss time was 6.3 +/- 6.7 years for prelingual loss and 12.1 +/- 11.6 years for postlingual. Mean follow-up was 3.9 +/- 2.8 years. Major complications occurred in 8.9% and minor in 7.8%. Problems during electrode insertion (3.8%) were the most frequent major complication followed by flap dehiscence (1.4%). Temporary facial palsy (2.2%), canal-wall lesion (2.2%) and tympanic membrane lesion (1.8%) were the more frequent minor complications. No death occurred. Conclusion: There was a low rate of surgical complications, most of them been successfully managed. These results confirm that cochlear implant is a safe surgery and most surgical complications can be managed with conservative measures or minimal intervention.-
dc.description.abstractA cirurgia de implante coclear (IC) é método seguro e factível de restaurar audição em pacientes com surdez grave a profunda. Objetivo: Descrever complicações do IC. Material e Métodos: Dados de 591 IC com tempo de seguimento de pelo menos 1 ano foram retrospectivamente analisados. Foram excluídos 41 pacientes por dados incompletos, acesso via fossa media ou perda de seguimento. Resultados: Dos 550 IC analisados, 341 foram realizados em crianças e adolescentes e 209 em adultos. O tempo médio de surdez foi de 6,3 ± 6,7 anos para pré-linguais e 12,1 ± 11,6 anos para os pós-linguais. O tempo médio de seguimento foi 3,9 ± 2,8 anos. Complicações maiores ocorreram em 8,9% casos e menores em 7,8%. Problemas durante a inserção do feixe de eletrodos (3,8%) foi a complicação maior mais frequente, seguido pela deiscência de ferida operatória (1,4%). Paralisia facial temporária (2,2%), lesão muro facial (2,2%) e lesão de membrana timpânica (1,8%) foram as complicações menores mais frequentes. Nenhum óbito foi observado. Conclusão: Complicações cirúrgicas tiveram baixa ocorrência, sendo a maioria solucionada com sucesso. Esses resultados confirmam que o IC é uma cirurgia segura e a maioria das complicações cirúrgicas pode ser manejada com medidas conservadoras ou intervenções menores.-
dc.language.isoeng-
dc.language.isopor-
dc.publisherASSOC BRASILEIRA OTORRINOLARINGOLOGIA & CIRURGIA CERVICOFACIAL-
dc.relation.ispartofBrazilian Journal of Otorhinolaryngology-
dc.rightsopenAccess-
dc.subjectcochlear implants-
dc.subjectdeafness-
dc.subjectintraoperative complications-
dc.subjectpostoperative complications-
dc.subjectComplicações intra-operatórias-
dc.subjectComplicações pós-operatórias-
dc.subjectImplante coclear-
dc.subjectSurdez-
dc.subject.othersurgery-
dc.subject.otherchildren-
dc.titleSurgical complications in 550 consecutive cochlear implantation-
dc.title.alternativeComplicações em 550 cirurgias consecutivas de implante coclear-
dc.typearticle-
dc.rights.holderCopyright ASSOC BRASILEIRA OTORRINOLARINGOLOGIA & CIRURGIA CERVICOFACIAL-
dc.identifier.doi10.1590/S1808-86942012000300014-
dc.identifier.pmid22714851-
dc.subject.wosOtorhinolaryngology-
dc.type.categoryoriginal article-
dc.type.versionpublishedVersion-
hcfmusp.description.beginpage80-
hcfmusp.description.endpage85-
hcfmusp.description.issue3-
hcfmusp.description.volume78-
hcfmusp.origemWOS-
hcfmusp.origem.id2-s2.0-84862904847-
hcfmusp.origem.idWOS:000305549100014-
hcfmusp.origem.idSCIELO:S1808-86942012000300014-
hcfmusp.publisher.citySAO PAULO-
hcfmusp.publisher.countryBRAZIL-
hcfmusp.relation.referenceAlbernaz PLM, 1996, REV BRAS MED OTORRIN, V3, P119-
hcfmusp.relation.referenceArnold W, 1997, AM J OTOL, V18, pS67-
hcfmusp.relation.referenceBento RF, 2001, ARQ INT OTORRINOLARI, V5, P130-
hcfmusp.relation.referenceBento RF, 2002, ARQ INT OTORRINOLARI, V6, P131-
hcfmusp.relation.referenceBibas A, 2006, Cochlear Implants Int, V7, P167, DOI 10.1002/cii.309-
hcfmusp.relation.referenceBilger RC, 1977, T SECT OTOLARYNGOL 1, V84-
hcfmusp.relation.referenceBrasil. Ministerio da Saude, 1999, DIARIO OFICIAL UNIAO-
hcfmusp.relation.referenceCBIC, CENTR BRAS IMPL COCL-
hcfmusp.relation.referenceCohen N L, 1988, Ann Otol Rhinol Laryngol Suppl, V135, P8-
hcfmusp.relation.referenceCOHEN NL, 1991, ANN OTO RHINOL LARYN, V100, P708-
hcfmusp.relation.referenceDing XY, 2009, ORL J OTO-RHINO-LARY, V71, P192, DOI 10.1159/000229297-
hcfmusp.relation.referenceGIBSON WPR, 1995, J LARYNGOL OTOL, V109, P821-
hcfmusp.relation.referenceHARRIS JP, 1987, LARYNGOSCOPE, V97, P755-
hcfmusp.relation.referenceHoffman RA, 1997, AM J OTOL, V18, P184-
hcfmusp.relation.referenceHoffman R. A., 1995, Annals of Otology Rhinology and Laryngology, V104, P420-
hcfmusp.relation.referenceHOUSE WF, 1982, ANN OTO RHINOL LARYN, V91, P15-
hcfmusp.relation.referenceJACKLER RK, 1987, LARYNGOSCOPE, V97, P801-
hcfmusp.relation.referenceKempf HG, 1999, LARYNGO RHINO OTOL, V78, P529-
hcfmusp.relation.referenceLima LRP, 2010, BRAZ J OTORHINOLAR, V76, P517-
hcfmusp.relation.referenceTelian SA, 1999, AM J OTOL, V20, P331-
hcfmusp.relation.referenceVenail F, 2008, ARCH OTOLARYNGOL, V134, P1276, DOI 10.1001/archoto.2008.504-
hcfmusp.relation.referenceWEBB RL, 1991, ANN OTO RHINOL LARYN, V100, P131-
hcfmusp.relation.referenceWootten CT, 2006, LARYNGOSCOPE, V116, P2055, DOI 10.1097/01.ndg.0000240286.43289.87-
dc.description.indexMEDLINE-
hcfmusp.lim.ref2012-
hcfmusp.citation.scopus47-
hcfmusp.scopus.lastupdate2024-03-08-
Appears in Collections:

Artigos e Materiais de Revistas Científicas - FM/MOF
Departamento de Otorrinolaringologia e Oftalmologia - FM/MOF

Artigos e Materiais de Revistas Científicas - HC/ICHC
Instituto Central - HC/ICHC

Artigos e Materiais de Revistas Científicas - LIM/32
LIM/32 - Laboratório de Otorrinolaringologia


Files in This Item:
File Description SizeFormat 
art_BRITO_Surgical_complications_in_550_consecutive_cochlear_implantation_2012_eng.PDFpublishedVersion (English)208.9 kBAdobe PDFThumbnail
View/Open
art_BRITO_Surgical_complications_in_550_consecutive_cochlear_implantation_2012_por.PDFpublishedVersion (Portuguese)221.62 kBAdobe PDFThumbnail
View/Open

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.