Modeling, characterization and test of a pediatric ventricular assist device

Carregando...
Imagem de Miniatura
Citações na Scopus
3
Tipo de produção
conferenceObject
Data de publicação
2018
Título da Revista
ISSN da Revista
Título do Volume
Editora
IOP PUBLISHING LTD
Autores
SOUSA, D. L.
CORDEIRO, T. D.
MELO, T. R.
ROCHA NETO, J. S. da
LIMA, A. M. N.
Citação
2017 JOINT IMEKO TC1-TC7-TC13 SYMPOSIUM: MEASUREMENT SCIENCE CHALLENGES IN NATURAL AND SOCIAL SCIENCES, v.1044, article ID UNSP 012047, 6p, 2018
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
This paper presents a OD model for the pulsatile pediatric Ventricular Assist Device (pVAD), as well as its characterization and test results obtained by using an experimental hydraulic system that provides operational conditions close to the physiological conditions observed in pediatric patients. The measurements of the physical quantities used in characterization of the pVAD have been obtained by using the LabVIEW platform. The air chamber pressure signal is used to model the pneumatic actuation system behaviour, whereas the pressure-volume relationship inside the blood chamber is applied to model the pVAD compliance. The set of test and validation results demonstrated the correctness of the proposed model.
Palavras-chave
Referências
  1. [Anonymous], TRUWAVE DISP PRESS T
  2. Benicio A, 1999, REV BRAS CIR CARDIOV, V14, P237
  3. em-tec GmbH, EXP NON FLOW MEAS
  4. FERRARI G, 1992, J BIOMED ENG, V14, P91, DOI 10.1016/0141-5425(92)90014-C
  5. Honeywell International Inc, PRESS SENS SCX SER
  6. Hunsberger A. Z, 2005, THESIS
  7. Olufsen MS, 2004, MATH BIOSCI ENG, V1, P61, DOI 10.3934/mbe.2004.1.61
  8. Pantalos GM, 2004, ASAIO J, V50, P37, DOI 10.1097/01.MAT.0000104818.70726.E6
  9. Shi YB, 2011, BIOMED ENG ONLINE, V10, DOI 10.1186/1475-925X-10-33
  10. Torres D.S, 2016, 25 CBEB